PRODUCTS

AV3672 series vector network analyzer include AV3672A (10MHz~13.5GHz), AV3672B (10MHz~26.5GHz), AV3672C (10MHz~40GHz), AV3672D (10MHz~50GHz) and AV3672E (10MHz~67GHz). In terms of hardware, comparing to last generation, the brand-new design concept and technical proposal have upgraded key performance indexes (KPI) of the complete machine such as sweep speed and dynamic range etc.; in terms of software, the embedded computer system (ECS) with high-performance microprocessor chip and the platform based on Windows operating system have greatly improved interconnectivity and usability of the complete machine. AV3672 series vector network analyzer provide multiple calibration types including frequency response, single port, response isolation, enhanced response and full dual-port, electronic calibration etc.; It offers various display formats such as logarithmic amplitude, linear amplitude, standing-wave, phase, group delay, Smith chart and polar coordinates, etc.; It is designed with several standard interfaces: USB, LAN, GPIB and VGA etc. Besides all measurement functions same as traditional vector network analyzer, through configuration of functional options, AV3672 series vector network analyzer is also capable of multifunctional & comprehensive parameter test of mixer/converter, gain compression two-dimensional sweep and pulse S-parameters, as well as accuracy measurement of amplitude-frequency characteristics, phase-frequency characteristics and group delay characteristics of microwave network. AV3672 series Vector Network Analyzer can be widely used in the field of transmitting/receiving (T/R) module measurement, dielectric material properties measurement, microwave pulse characteristics measurement and optoelectronic properties measurement, which is an indispensable instrument for scientific research and manufacturing process of radar, communication, and navigation systems. 
Ø Flexible and optional calibration types, compatible with multiple calibration kits
Ø Support multi-window, multi-channel tests, fast implementation of complex test scheme
Ø Available in multiple display formats such as logarithmic amplitude, linear amplitude, standing-wave, phase, Smith chart
Ø With USB, GPIB, LAN and VGA display interface
Ø Chinese/English operation interface, 12.1-inch high resolution touch screen
Ø Record/run, one-click operation greatly simplifies the measurement setting steps and improves the efficiency
Ø Available in functions as pulse S parameter measurement, time-domain measurement, mixer measurement, gain compression two-dimensional sweep measurement, millimeter-wave frequency extension, antenna & RCS measurement and receiving etc.


Humanized user interface for easy operation, which can improve the efficiency 



Flexible and optional calibration types, compatible with multiple calibration kits
AV3672 series vector network analyzer provides multiple calibration types, including guided calibration (automatic calibration), unguided calibration (using mechanical calibration kit to conduct through response calibration, through response & isolation calibration, single port calibration, enhanced response calibration, full two-port SOLT calibration, TRL calibration) and electronic calibration (ECal) etc. Users can select calibration kits, such as coaxial 3.5mm calibration kit and electronic calibration kit based on test requirements, which greatly facilitates testing on devices with different interfaces.


Multiple windows to display all measuring channels
The analyzer has function of multi-channel and multi-window display. It supports up to 64 channels. Maximum 16 measuring windows can be simultaneously displayed, and each window can simultaneously display up to 8 testing traces, which makes the observation results more visible and the operation more convenient. 



Chinese/English operation interface, 12.1-inch high resolution touch screen
AV3672 series vector network analyzer adopts Chinese/English operation interface. The 12.1-inch touch screen with 1280*800 resolution has bright and comfortable color, which can make your operation very convenient. 

Large dynamic range
AV3672 series vector network analyzer is designed with the concept of mixer receiving, which effectively extends the dynamic range of the complete machine and meets the test demand for large dynamic range. 


Rich peripheral interfaces, flexible and practical
Adopting the software & hardware platform consisting of embedded computer module compatible with PC and Windows operation system, AV3672 series vector network analyzer realizes the perfect combination of the instrument and PC. Users can use the rich I/O interfaces (including GPIB, USB, and LAN etc.) to complete the optimum selection of data communication. 
 

Low trace noise, high test accuracy
The excellent performance of AV3672 series vector network analyzer in trace noise highly improves its test accuracy so as to meet user’s demand for accurate measurement, and it is especially helpful for the accurate measurement of devices with low insertion loss. (AV3672B is taken as an example below)


Time-domain analysis can comprehensively characterize the design
With time-domain options, AV3672 series vector network analyzer can realize the switching of measurement results between frequency-domain and time-domain, which can be used to identify the discontinuous points of devices, fixtures or cables to realize accurate fault location. 

Automatic test
Calibration is complicated: short-circuiter, open-circuiter, load, through – SO PROFESSIONAL!
Set up the same parameters over and over again – VERY BORING!
Products need to be tested under harsh conditions – But PEOPLE needn’t!
Setting changing and data recording need to be done at set intervals – Unfortunately,I am not a Timer!


AV3672 series vector network analyzer can provide you integrated automatic test solutions including automatic calibration, automatic measurement, automatic reading and automatic printing.
Flexible and multiple control modes are provided through GPIB, LAN, and USB interfaces. All you need to do is to finish the interconnection of equipment and send the command.



Mixer test
The 4-port measurement option of AV3672 series vector network analyzer has two built-in sources, which can finish the test of scalar and vector parameters of mixers

 
Filter test
AV3672 series vector network analyzer has filter test menu, which is convenient and efficient.

 

Integrated pulse S parameter test
AV3672 series vector network analyzer has function of pulse modulation signal output and can conduct the test of pulse S parameters.

 
High-speed sweep magnetron test
AV3672 series vector network analyzer has high sweep speed, and so is capable of magnetron test.

 


AV3672A/B Technical Specifications
Frequency Characteristic  
Frequency Range 10MHz~13.5/26.5GHz
Frequency Resolution 1Hz
Frequency Accuracy ±1×10-7(23℃±3℃)
Port Harmonic Suppression Typical Value
Harmonic Suppression on port 1 and 3 -48dBc(0.01~4GHz)
-57dBc(4~13.5GHz)
-57dBc(13.5~26.5GHz)
-51dBc(0.01~4GHz)
-60dBc(4~13.5GHz)
-60dBc(13.5~26.5GHz)
Harmonic Suppression on port 2 and 4 -13dBc(0.01~4GHz)
-18dBc(4~13.5GHz)
-18dBc(13.5~26.5GHz)
-13dBc(0.01~4GHz)
-21dBc(4~13.5GHz)
-21dBc(13.5~26.5GHz)
Port Power Characteristic Typical Value
Power Sweep Range 33dB(10~500MHz)
30dB(0.5~4GHz)
34dB(4~7GHz)
31dB(7~13.5GHz)
29dB(13.5~20GHz)
25dB(20~26.5GHz)
42dB(10~500MHz)
37dB(0.5~4GHz)
42dB(4~7GHz)
41dB(7~13.5GHz)
38dB(13.5~20GHz)
35dB(20~26.5GHz)
Max. Output Power
(standard configuration, option 400)
Port1、3 0dBm(0.01~4GHz)(Filter mode)
+9dBm(10~50MHz)(High power mode)
+6dBm(0.05~4GHz)(High power mode)
+12dBm(4~7GHz)
+8dBm(7~13.5GHz)
+6dBm(13.5~20GHz)
+4dBm(20~26.5GHz)
  +16dBm(10~50MHz)(High power mode)
+11dBm(0.05~4GHz)(High power mode)
+16dBm(4~7GHz)
+14dBm(7~13.5GHz)
+11dBm(13.5~20GHz)
+9dBm(20~26.5GHz)
Port2、4 +13dBm(10~50MHz)
+13dBm(0.05~4GHz)
+10dBm(4~7GHz)
+9dBm(7~13.5GHz)
+6dBm(13.5~20GHz)
+2dBm(20~26.5GHz)
+16dBm(10~50MHz)
+16dBm(0.05~4GHz)
+16dBm(4~7GHz)
+15dBm(7~13.5GHz)
+13dBm(13.5~20GHz)
+8dBm(20~26.5GHz)
Max. Output Power
(option 201, 401, 402)
Port1、3 -1dBm(0.01~4GHz)(Filter mode)
+8dBm(10~50MHz)(High power mode)
+5dBm(0.05~4GHz)(High power mode)
+10dBm(4~7GHz)
+6dBm(7~13.5GHz)
+4dBm(13.5~20GHz)
+2dBm(20~26.5GHz)
+15dBm(10~50MHz)(High power mode)
+10dBm(0.05~4GHz)(High power mode)
+15dBm(4~7GHz)
+13dBm(7~13.5GHz)
+10dBm(13.5~20GHz)
+8dBm(20~26.5GHz)
Port2、4 +12dBm(10~50MHz)
+12dBm(0.05~4GHz)
+9dBm(4~7GHz)
+8dBm(7~13.5GHz)
+5dBm(13.5~20GHz)
0dBm(20~26.5GHz)
+15dBm(10~50MHz)
+15dBm(0.05~4GHz)
+15dBm(4~7GHz)
+14dBm(7~13.5GHz)
+12dBm(13.5~20GHz)
+7dBm(20~26.5GHz)
1dB Compression Level / +13dBm
Pulse Characteristic Typical Value
Pulse Width Setting Range 33ns~60s 20ns~60s
Pulse Transition Time
(10%-90%)
30ns 20ns
Pulse on/off Ratio 64dB(0.01~4GHz)
80dB(4~26.5GHz)
Network Parameter Characteristic Typical Value
System Dynamic Range 90dB(0.01~1GHz)
120dB(1~4GHz)
127dB(4~10GHz)
120dB(10~20GHz)
115dB(20~24GHz)
110dB(24~26.5GHz)
100dB(0.01~1GHz)
135dB(1~4GHz)
135dB(4~10GHz)
132dB(10~20GHz)
129dB(20~24GHz)
125dB(24~26.5GHz)
Effective Directivity 48dB(0.01~2GHz)
44dB(2~26.5GHz)
60dB(0.01~2GHz)
53dB(2~26.5GHz)
Effective Source Match 40dB(0.01~2GHz)
30dB(2~26.5GHz)
46dB(0.01~2GHz)
36dB(2~26.5GHz)
Effective Load Match 48dB(0.01~2GHz)
44dB(2~26.5GHz)
60dB(0.01~2GHz)
51dB(2~26.5GHz)
Reflection Tracking ±0.04dB(0.01~2GHz)
±0.05dB(2~26.5GHz)
±0.004dB(0.01~2GHz)
±0.010dB(2~26.5GHz)
Transmission Tracking ±0.10dB(0.01~2GHz)
±0.12dB(2~26.5GHz)
±0.005dB(0.01~2GHz)
±0.015dB(2~26.5GHz)
Others Typical Value
Amplitude Trace Noise
dB rms(1kHz IF bandwidth)
0.0500(10~50MHz)
0.0070(50~500MHz)
0.0020(0.5~13.5GHz)
0.0030(13.5~20GHz)
0.0050(20~26.5GHz)
0.0020(10~50MHz)
0.0008(50~500MHz)
0.0009(0.05~13.5GHz)
0.0008(13.5~20GHz)
0.0010(20~26.5GHz)
Phase Trace Noise
deg rms(1kHz IF bandwidth)
0.200(10~50MHz)
0.051(50~500MHz)
0.042(0.5~13.5GHz)
0.054(13.5~20GHz)
0.054(20~26.5GHz)
0.020(10~50MHz)
0.006(50~500MHz)
0.006(0.05~13.5GHz)
0.006(13.5~20GHz)
0.012(20~26.5GHz)
IF Bandwidth 1Hz~5MHz
Amplitude Display Resolution 0.001dB/div
Phase Display Resolution 0.01°/div
Setting Requirement of Reference Level Amplitude -500~+500dB
Setting Requirement of Reference Level Phase -500~+500°
General Characteristic
Port Connectors 3.5mm (male), 50 ohm system impedance
Number of Measuring Ports   AV3672A/B standard configuration:2 ports;AV3672A/B-400 option:4 ports
Peripheral Interface USB、GPIB、VGA、LAN
Operating System Windows 7
Display 12.1-inch high resolution touch screen
Dimensions W´H´D=426mm×266mm×550mm (excluding bottom foot and handle)
W´H´D=516mm×280mm×640mm (including handle, bottom foot and back foot)
Max. Power Consumption 400W
Max. Weight 42kg

AV3672C/D Technical Specifications
Frequency Characteristic
Frequency Range 10MHz~43.5/50GHz
Frequency Resolution 1Hz
Frequency Accuracy ±1×10-7(23℃±3℃)
Port Harmonic Suppression Typical Value
Harmonic Suppression on Port 1 and 3 -48dBc(0.01~4GHz)
-57dBc(4~13.5GHz)
-57dBc(13.5~50GHz)
-51dBc(0.01~4GHz)
-60dBc(4~13.5GHz)
-60dBc(13.5~50GHz)
Harmonic Suppression on Port 2 and 4 -13dBc(0.01~4GHz)
-18dBc(4~13.5GHz)
-57dBc(13.5~50GHz)
-13dBc(0.01~4GHz)
-21dBc(4~13.5GHz)
-60dBc(13.5~50GHz)
Port Power Characteristic Typical Value
Power Sweep Range 32dB(10~50MHz)
29dB(0.05~4GHz)
28dB(4~13.5GHz)
30dB(13.5~40GHz)
27dB(40~47GHz)
15dB(47~50GHz)
41dB(10~500MHz)
40dB(0.5~4GHz)
36dB(4~13.5GHz)
38dB(13.5~40GHz)
36dB(40~47GHz)
23dB(47~50GHz)
Max. Output Power
(standard configuration, option 400)
Port 1、3 -1dBm(10~50MHz)(Filter mode)
0dBm(0.05~4GHz)(Filter mode)
+8dBm(10~50MHz)(High power mode)
+5dBm(0.05 ~4GHz)(High power mode)
+5dBm(4~13.5GHz)
+7dBm(13.5~40GHz)
+5dBm(40~47GHz)
-7dBm(47~50GHz)
+13dBm(10~50MHz)(High power mode)
+9dBm(0.05~4GHz)(High power mode)
+10dBm(4~13.5GHz)
+13dBm(13.5~40GHz)
+10dBm(40~47GHz)
0dBm(47~50GHz)
Port 2、4 +11dBm(10~50MHz)
+9dBm(0.05~4GHz)
+6dBm(4~13.5GHz)
+7dBm(13.5~40GHz)
+5dBm(40~47GHz)
-7dBm(47~50GHz)
+16dBm(10~50MHz)
+15dBm(0.05~4GHz)
+13dBm(4~13.5GHz)
+12dBm(13.5~40GHz)
+9dBm(40~47GHz)
-1dBm(47~50GHz)
Max. Output Power
(option 201, 401, 402)
Port 1、3 -2dBm(10~50MHz)(Filter mode)
-1dBm(0.05~4GHz)(Filter mode)
+7dBm(10~50MHz)(High power mode)
+4dBm(0.05~4GHz)(High power mode)
+3dBm(4~13.5GHz)
+5dBm(13.5~40GHz)
+2dBm(40~47GHz)
-10dBm(47~50GHz)
+12dBm(10~50MHz)(High power mode)
+8dBm(0.05~4GHz)(High power mode)
+9dBm(4~13.5GHz)
+12dBm(13.5~40GHz)
+9dBm(40~47GHz)
-1dBm(47~50GHz)
Port 2、4 +10dBm(10~50MHz)
+8dBm(0.05~4GHz)
+4dBm(4~13.5GHz)
+5dBm(13.5~40GHz)
+2dBm(40~47GHz)
-10dBm(47~50GHz)
+15dBm(10~50MHz)
+14dBm(0.05~4GHz)
+12dBm(4~13.5GHz)
+11dBm(13.5~40GHz)
+8dBm(40~47GHz)
-2dBm(47~50GHz)
1dB Compression Level / +10dBm
Pulse Characteristic Typical Value
Pulse Width Setting Range 33ns~60s 20ns~60s
Pulse Transition Time
(10%-90%)
30ns 20ns
Pulse on/off Ratio 64dB(0.01~4GHz)
80dB(4~40GHz)
80dB(40~50GHz)
Network Parameter Characteristic Typical Value
System Dynamic Range 74dB(0.01~1GHz)
119dB(1~13.5GHz)
115dB(13.5~26.5GHz)
110dB(26.5~35GHz)
105dB(35~47GHz)
90dB(47~50GHz)
105dB(0.01~1GHz)
133dB(1~13.5GHz)
126dB(13.5~26.5GHz)
120dB(26.5~35GHz)
116dB(35~47GHz)
103dB(47~50GHz)
Effective Directivity 42dB(0.01~13.5GHz)
38dB(13.5~40GHz)
36dB(40~50GHz)
50dB(0.01~13.5GHz)
45dB(13.5~40GHz)
42dB(40~50GHz)
Effective Source Match 36dB(0.01~2GHz)
31dB(2~13.5GHz)
28dB(13.5~40GHz)
27dB(40~50GHz)
43dB(0.01~2GHz)
34dB(2~13.5GHz)
33dB(13.5~40GHz)
33dB(40~50GHz)
Effective Load Match 42dB(0.01~13.5GHz)
37dB(13.5~40GHz)
35dB(40~50GHz)
60dB(0.01~13.5GHz)
56dB(13.5~40GHz)
51dB(40~50GHz)
Reflection Tracking ±0.04dB(0.01~13.5GHz)
±0.03dB(13.5~40GHz)
±0.04dB(40~50GHz)
±0.010dB(0.01~13.5GHz)
±0.010dB(13.5~40GHz)
±0.020dB(40~50GHz)
Transmission Tracking ±0.1dB(0.01~13.5GHz)
±0.16dB(13.5~40GHz)
±0.20dB(40~50GHz)
±0.012dB(0.01~13.5GHz)
±0.015dB(13.5~40GHz)
±0.020dB(40~50GHz)
Others Typical Value
Amplitude Trace Noise
dB rms(1kHz IF bandwidth)
0.050(10~50MHz)
0.020(50~500MHz)
0.005(0.5~13.5GHz)
0.004(13.5~26.5GHz)
0.008(26.5~50GHz)
0.0060(10~50MHz)
0.0020(50~500MHz)
0.0010(0.5~13.5GHz)
0.0009(13.5~26.5GHz)
0.0040(26.5~50GHz)
Phase Trace Noise
deg rms (1kHz IF bandwidth)
0.900(10~50MHz)
0.700(50~500MHz)
0.040(0.5~13.5GHz)
0.050(13.5~26.5GHz)
0.060(26.5~50GHz)
0.040(10~50MHz)
0.010(50~500MHz)
0.005(0.5~13.5GHz)
0.020(13.5~26.5GHz)
0.030(26.5~50GHz)
IF Bandwidth 1Hz~5MHz
Amplitude Display Resolution 0.001dB/div
Phase Display Resolution 0.01°/div
Setting Requirement of Reference Level Amplitude -500~+500dB
Setting Requirement of Reference Level Phase -500~+500°
General Characteristic  
Port Connectors 2.4mm (male), 50 ohm system impedance
Number of measuring Ports AV3672C/D standard configuration:2 ports, AV3672C/D-400 option: 4 ports
Peripheral Interface USB, GPIB, VGA, LAN
Operating System Windows 7
Display 12.1-inch high resolution touch screen
Dimensions W´H´D=426mm×266mm×600mm (excluding bottom foot and handle)
W´H´D=516mm×280mm×690mm (including handle, bottom foot and back foot)
Max. Power Consumption 500W
Max. Weight 47kg
 

AV3672E Technical Specifications
Frequency Characteristics  
Frequency Range 10MHz~67GHz
Frequency Resolution 1Hz
Frequency Accuracy ±1×10-7(23℃±3℃)
Port Harmonic Suppression Typical Value
Harmonic Suppression on Port 1 and 3 -48dBc(0.01~4GHz)
-57dBc(4~67GHz)
-51dBc(0.01~4GHz)
-60dBc(4~67GHz)
Harmonic Suppression on Port 2 and 4 -13dBc(0.01~4GHz)
-18dBc(4~13.5GHz)
-57dBc(13.5~67GHz)
-13dBc(0.01~4GHz)
-21dBc(4~13.5GHz)
-60dBc(13.5~67GHz)
Port Power Characteristics Typical Value
Power Sweep Range 32dB(10~50MHz)
29dB(0.05~4GHz)
28dB(4~26.5GHz)
29dB(26.5~35GHz)
26dB(35~40GHz)
25dB(40~67GHz)
42dB(10~500MHz)
38dB(0.5~4GHz)
36dB(4~26.5GHz)
35dB(26.5~35GHz)
34dB(35~40GHz)
32dB(40~67GHz)
Max. Output Power
(standard configuration, option 400)
Port 1、3 -1dBm(10~50MHz)(Filter mode)
0dBm(0.05~4GHz)(Filter mode)
+8dBm(10~50MHz)(High power mode)
+5dBm(0.05~4GHz)(High power mode)
+1dBm(4~13.5GHz)
+5dBm(13.5~26.5GHz)
+3dBm(26.5~40GHz)
+5dBm(40~67GHz)
+16dBm(10~50MHz)(High power mode)
+10dBm(0.05~4GHz)(High power mode)
+9dBm(4~13.5GHz)
+11dBm(13.5~26.5GHz)
+10dBm(26.5~40GHz)
+9dBm(40~67GHz)
Port 2、4 +8dBm(10~50MHz)
+5dBm(0.05~4GHz)
+1dBm(4~13.5GHz)
+5dBm(13.5~26.5GHz)
+3dBm(26.5~40GHz)
+5dBm(40~67GHz)
+16dBm(10~50MHz)
+15dBm(0.05~4GHz)
+10dBm(4~13.5GHz)
+11dBm(13.5~26.5GHz)
+9dBm(26.5~40GHz)
+8dBm(40~67GHz)
Max. Output Power
(option 201, 401, 402)
Port 1、3 -2dBm(10~50MHz)(Filter mode)
-1dBm(0.05~4GHz)(Filter mode)
+7dBm(10~50MHz)(High power mode)
+4dBm(0.05~4GHz)(High power mode)
-2dBm(4~13.5GHz)
+3dBm(13.5~26.5GHz)
0dBm(26.5~67GHz)
+15dBm(10~50MHz)(High power mode)
+9dBm(0.05~4GHz)(High power mode)
+6dBm(4~13.5GHz)
+7dBm(13.5~26.5GHz)
+4dBm(26.5~67GHz)
Port 2、4 +7dBm(10~50MHz)
+4dBm(0.05~4GHz)
-2dBm(4~13.5GHz)
+3dBm(13.5~26.5GHz)
0dBm(26.5~67GHz)
+15dBm(10~50MHz)
+14dBm(0.05~4GHz)
+9dBm(4~13.5GHz)
+10dBm(13.5~26.5GHz)
+5dBm(26.5~67GHz)
1dB Compression Level / +10dBm
Pulse Characteristics Typical Value
Pulse Width Setting Range 33ns~60s 20ns~60s
Pulse Transition Time (10%-90%) 30ns 20ns
Pulse on/off Ratio 64dB(0.01~4GHz)
80dB(4~67GHz)
Network Parameter Characteristic Typical Value
System Dynamic Range 74dB(0.01~1GHz)
100dB(1~4GHz)
120dB(4~10GHz)
112dB(10~26.5GHz)
108dB(26.5~35GHz)
105dB(35~50GHz)
100dB(50~67GHz)
100dB(0.01~1GHz)
125dB(1~4GHz)
125dB(4~10GHz)
120dB(10~26.5GHz)
115dB(26.5~35GHz)
112dB(35~50GHz)
105dB(50~67GHz)
Effective Directivity 35dB(0.01~2GHz)
41dB(2~13.5GHz)
34dB(13.5~40GHz)
32dB(40~67GHz)
50dB(0.01~2GHz)
50dB(2~13.5GHz)
50dB(13.5~40GHz)
42dB(40~67GHz)
Effective Load Match 35dB(0.01~2GHz)
41dB(2~13.5GHz)
33dB(13.5~40GHz)
30dB(40~67GHz)
60dB(0.01~2GHz)
50dB(2~13.5GHz)
50dB(13.5~40GHz)
45dB(40~67GHz)
Reflection Tracking ±0.05dB(0.01~2GHz)
±0.06dB(2~13.5GHz)
±0.08dB(13.5~40GHz)
±0.10dB(40~67GHz)
±0.005dB(0.01~2GHz)
±0.005dB(2~13.5GHz)
±0.008dB(13.5~40GHz)
±0.010dB(40~67GHz)
Transmission Tracking ±0.10dB(0.01~2GHz)
±0.11dB(2~13.5GHz)
±0.16dB(13.5~40GHz)
±0.20dB(40~67GHz)
±0.005dB(0.01~2GHz)
±0.006dB(2~13.5GHz)
±0.015dB(13.5~40GHz)
±0.020dB(40~67GHz)
Others   Typical Value
Amplitude Trace Noise
dB rms(1kHz IF bandwidth)
0.050(10~50MHz)
0.020(50~500MHz)
0.005(0.5~13.5GHz)
0.004(13.5~26.5GHz)
0.020(26.5~67GHz)
0.0090(10~50MHz)
0.0020(50~500MHz)
0.0008(0.5~13.5GHz)
0.0008(13.5~26.5GHz)
0.0050(26.5~67GHz)
Phase Trace Noise
deg rms(1kHz IF bandwidth)
0.90(10~50MHz)
0.70(50~500MHz)
0.04(0.5~13.5GHz)
0.05(13.5~26.5GHz)
0.10(26.5~67GHz)
0.010(10~50MHz)
0.010(50~500MHz)
0.006(0.5~13.5GHz)
0.007(13.5~26.5GHz)
0.030(26.5~67GHz)
IF Bandwidth 1Hz~5MHz
Amplitude Display Resolution 0.001dB/div
Phase Display Resolution 0.01°/div
Setting Requirement of Reference Level Amplitude -500~+500dB
Setting Requirement of Reference Level Phase -500~+500°
General Characteristics  
Port Connectors 1.85mm (male), 50 ohm system impedance
Number of measuring Ports AV3672E standard configuration:2 ports, AV3672E-400 option: 4 ports
Peripheral Interface USB, GPIB, VGA, LAN
Operating System Windows 7
Display 12.1-inch high resolution touch screen
Dimensions W´H´D=426mm×266mm×600mm (excluding bottom foot and handle)
W´H´D=516mm×280mm×690mm (including handle, bottom foot and back foot)
Max. Power Consumption 500W
Max. Weight 50 kg

 

Main Unit Description
AV3672A Vector network analyzer (10MHz ~ 13.5GHz)
AV3672B Vector network analyzer (10MHz ~ 26.5GHz)
AV3672C Vector network analyzer (10MHz ~ 40GHz)
AV3672D Vector network analyzer (10MHz ~ 50GHz)
AV3672E Vector network analyzer (10MHz ~ 67GHz)
 
Standard Configuration
No. Description Remarks
1 Power cord assembly Standard three-core power cord
2 USB keyboard/mouse  
3 User manual  
4 Certificate of conformity  
5 Aluminum alloy box  
 
AV3672A Options
Model Description Remarks
AV3672A-201 2-port programmable step attenuator Set two 70dB programmable step attenuators for the source path, and two 35dB programmable step attenuators for the receiver path
AV3672A-400 4-port measurement Two-source stimulus four-port VNA configuration
AV3672A-401 4-port programmable step attenuator Set four 70dB programmable step attenuators for the source path, and four 35dB programmable step attenuators for the receiver path (option 400 is needed)
AV3672B-402 Active intermodulation measurement For intermodulation signal measurement of amplifier (option 400 is needed)
AV3672B-006 English options Key, front panel, label, operation system
AV3672A-008 Pulse measurement For pulse S-parameter measurement
AV3672A-S10 Time domain measurement For time-domain test, can locate and analyze the discontinuous positions in devices, fixtures or cables.
AV3672A-S80 Frequency offset measurement For frequency offset measurement. Millimeter-wave frequency extension main unit needs this option
AV3672A-S82 Mixer scalar measurement For the scalar parameter measurement of mixers (option 400 is needed)
AV3672A-S83 Mixer vector measurement For the vector parameter measurement of mixers (option 400 is needed)
AV3672A-S84 Embedded LO frequency   converter measurement For the measurement of embedded LO frequency converters (option 400, S82 or S83 are needed)
AV3672A-S86 Gain compression two-dimension sweep measurement For the gain compression two-dimension sweep test of amplifier
AV31121 3.5mm calibration kit For calibration of the analyzer
FB0HA0HB025.0 3.5mm testing cable For measurement of the analyzer
FB0HA0HC025.0 3.5mm testing cable For measurement of the analyzer
AV20403 E-cal kit For calibration of the analyzer (10MHz-26.5GHz, 2 ports)
AV20405 E-cal kit For calibration of the analyzer (10MHz-20GHz, 4 ports)
AV87232 USB Power Probe For 402, S82, S86 options in the process of power calibration(50MHz-26.5GHz)
AV3672B Options
Model Description Remarks
AV3672B-201 2-port programmable step attenuator Set two 70dB programmable step attenuators for the source path, and two 35dB programmable step attenuators for the receiver path
AV3672B-400 4-port measurement Two-source stimulus four-port VNA configuration
AV3672B-401 4-port programmable step attenuator Set four 70dB programmable step attenuators for the source path, and four 35dB programmable step attenuators for the receiver path (option 400 is needed)
AV3672B-402 Active intermodulation measurement For intermodulation signal measurement of amplifier (option 400 is needed)
AV3672B-006 English options Key, front panel, label, operation system
AV3672B-008 Pulse measurement For pulse S-parameter measurement
AV3672B-S10 Time domain measurement For time-domain test, can locate and analyze the discontinuous positions in devices, fixtures or cables.
AV3672B-S80 Frequency offset measurement For frequency offset measurement. Millimeter-wave frequency extension main unit needs this option
AV3672B-S82 Mixer scalar measurement For the scalar parameter measurement of mixers (option 400 is needed)
AV3672B-S83 Mixer vector measurement For the vector parameter measurement of mixers (option 400 is needed)
AV3672B-S84 Embedded LO frequency converter measurement For the measurement of embedded LO frequency converters(option 400, S82 or S83 are needed)
AV3672B-S86 Gain compression two-dimension sweep measurement For the gain compression two-dimension sweep test of amplifier
AV31121 3.5mm calibration kit For calibration of the analyzer
FB0HA0HB025.0 3.5mm testing cable For measurement of the analyzer
FB0HA0HC025.0 3.5mm testing cable For measurement of the analyzer
AV20403 E-cal kit For calibration of the analyzer (10MHz-26.5GHz, 2 ports)
AV20405 E-cal kit For calibration of the analyzer (10MHz-20GHz, 4 ports)
AV87232 USB Power Probe For 402, S82, S86 options in the process of power calibration(50MHz-26.5GHz)



V3672C Options
Model Description Remarks
AV3672C-201 2-port programmable step attenuator Set two 60dB programmable step attenuators for the source path, and two 35dB programmable step attenuators for the receiver path
AV3672C-400 4-port measurement Two-source stimulus four-port VNA configuration
AV3672C-401 4-port programmable step attenuator Set four 60dB programmable step attenuators for the source path, and four 35dB programmable step attenuators for the receiver path (option 400 is needed)
AV3672B-402 Active intermodulation measurement For intermodulation signal measurement of amplifier (option 400 is needed)
AV3672B-006 English options Key, front panel, label, operation system
AV3672C-008 Pulse measurement For pulse S-parameter measurement
AV3672C-S10 Time domain measurement For time-domain test, can locate and analyze the discontinuous positions in devices, fixtures or cables.
AV3672C-S80 Frequency offset measurement For frequency offset measurement. Millimeter-wave frequency extension main unit needs this option
AV3672B-S82 Mixer scalar measurement For the scalar parameter measurement of mixers (option 400 is needed)
AV3672B-S83 Mixer vector measurement For the vector parameter measurement of mixers (option 400 is needed)
AV3672B-S84 Embedded LO frequency converter measurement For the measurement of embedded LO frequency converters(option 400, S82 or S83 are needed)
AV3672C-S86 Gain compression two-dimension sweep measurement For the gain compression two-dimension sweep test of amplifier
AV31123 2.4mm calibration kit For calibration of the analyzer
FE0BN0BM025.0 2.4mm testing cable For measurement of the analyzer
FE0BN0BL025.0 2.4mm testing cable For measurement of the analyzer
AV20404 E-cal kit For calibration of the analyzer (10MHz-50GHz 2 ports)
AV87233 USB Power Probe For 402, S82, S86 options in the process of power calibration(50MHz-40GHz)
AV3672D Options
Model Description Remarks
AV3672D-201 2-port programmable step attenuator Set two 60dB programmable step attenuators for the source path, and two 35dB programmable step attenuators for the receiver path
AV3672D-400 4-port measurement Two-source stimulus four-port VNA configuration
AV3672D-401 4-port programmable step attenuator Set four 60dB programmable step attenuators for the source path, and four 35dB programmable step attenuators for the receiver path (option 400 is needed)
AV3672B-006 English options Key, front panel, label, operation system
AV3672D-008 Pulse measurement For pulse S-parameter measurement
AV3672D-S10 Time domain measurement For time-domain test, can locate and analyze the discontinuous positions in devices, fixtures or cables.
AV3672D-S80 Frequency offset measurement For frequency offset measurement. Millimeter-wave frequency extension main unit needs this option
AV3672B-S82 Mixer scalar measurement For the scalar parameter measurement of mixers (option 400 is needed)
AV3672B-S83 Mixer vector measurement For the vector parameter measurement of mixers (option 400 is needed)
AV3672B-S84 Embedded LO frequency converter measurement For the measurement of embedded LO frequency converters(option 400, S82 or S83 are needed)
AV3672D-S86 Gain compression two-dimension sweep measurement For the gain compression two-dimension sweep test of amplifier
AV31123A 2.4mm calibration kit For calibration of the analyzer
FE0BN0BM025.0 2.4mm testing cable For measurement of the analyzer
FE0BN0BL025.0 2.4mm testing cable For measurement of the analyzer
AV20404 E-cal kit For calibration of the analyzer (10MHz-50GHz 2 ports)

    
AV3672E options:
Model Description Remarks
AV3672E-201 2-port programmable step attenuator Set two 50dB programmable step attenuators for the source path, and two 50dB programmable step attenuators for the receiver path
AV3672E-400 4-port measurement Two-source stimulus four-port VNA configuration
AV3672E-401 4-port programmable step attenuator Set four 50dB programmable step attenuators for the source path, and four 50dB programmable step attenuators for the receiver path (option 400 is needed)
AV3672E-402 Active intermodulation measurement For intermodulation signal measurement of amplifier (option 400 is needed)
AV3672B-006 English options Key, front panel, label, operation system
AV3672E-008 Pulse measurement For pulse S-parameter measurement
AV3672E-S10 Time-domain measurement For time-domain test, can locate and analyze the discontinuous positions in devices, fixtures or cables.
AV3672E-S80 Frequency offset measurement For frequency offset measurement. Millimeter-wave frequency extension main unit needs this option
AV3672E-S82 Mixer scalar measurement For the scalar parameter measurement of mixers (option 400 is needed)
AV3672E-S83 Mixer vector measurement For the vector parameter measurement of mixers (option 400 is needed)
AV3672E-S84 Embedded LO frequency converter measurement For the measurement of embedded LO frequency converters(option 400, S82 or S83 are needed)
AV3672E-S86 Gain compression two-dimension sweep measurement For the gain compression two-dimension sweep test of amplifier
AV31128 1.85mm calibration kit For calibration of the analyzer
N4697F/ FF0CN0CM025.0
FF0CN0CL025.0
1.85mm testing cable For measurement of the analyzer
AV20409 E-cal kit For calibration of the analyzer (10MHz-67GHz 2 ports)
 
Name:
Tel:
*Email:
Content: